Control Characteristics of pH-Controller
Home
Theory
Procedure
video
Simulator
Self Evaluation
References
Self-Evaluation
Q1. The desirable feature of measurement by any instrument is-
(A)
Quick response
(B)
Reproducible response
(C)
Stable response
(D)
All of the above
Q2. The essential desirable feature of any controller is -
(A)
Quick establishment of control action with-out any offset between the measured value and set point
(B)
Quick establishment of control action with-out any oscillations between the measured value and set point
(C)
Quick establishment of control action with-out any offset & oscillations between the measured value and set point
(D)
Quick establishment of control action with or with-out any offset & oscillations between the measured value and set point
Q3. The decrease in pH in any fermentation is primarily due to-
(A)
Consumption of carbon source
(B)
Consumption of ammonia (Nitrogen) source
(C)
Consumption of vitamins and growth factors
(D)
Accumulation of metabolites
Q4. Adjustment of “p-band” to 0.5 and set point of pH as 5 will mean -
(A)
The pH will be controlled in the range 4.5 to 5.5
(B)
The pH will be controlled in the range 4.5 to 5.0
(C)
The pH will be controlled in the range 4.0 to 5.5
(D)
The pH will be controlled in the range 4.0 to 5.0
Q5. The pH control action in the bioreactor is dependent on -
(A)
Density and volume of the fermentation broth
(B)
Strength of acid /bases used for the control
(C)
Addition of amount of acid /bases & pH range of addition in the bioreactor
(D)
All of the above
Q6. The td value of pH controller panel of bio-engineering fermenter signifies -
(A)
Safety time for the process after expiry of which if no control action takes place then the controller will give an alarm.
(B)
Dead time during which controller will not work at all
(C)
Dead time during which no acid and alkali addition will take place
(D)
None of the above is true
Q7. The pH meter has a temperature measurement device for -
(A)
Good record keeping
(B)
Temperature compensation of the pH readings
(C)
Both 1& 2 are correct
(D)
None of the above statements are true
Q8. Low value of p-band setting of the pH controller of Bioengineering fermenter gives
(A)
Quicker but oscillatory response of control action
(B)
Offset between controlled value of pH and the set point
(C)
Unstability of the control action as the controller action will trigger with minor change in pH values
(D)
None of the above
Q9. High value of p-band setting of the pH controller of Bio-engineering fermenter features -
(A)
Offset between controlled value of pH and the set point
(B)
Oscillatory response of control action
(C)
Larger time of control action
(D)
Insensitivity of controller in a wider pH range and improper control action
Q10. The pH probe calibration is done by -
(A)
At least two buffers to have a good linearity between the pH range of measurement
(B)
One buffer near the set point pH of the bioreactor
(C)
Either of above (1 or 2)
(D)
Neither of above (1 or 2)